GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 6
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8.) Unscrew 6 screws to remove 17" LCD front cover. Refer to the figure below.
9.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-1-3
Mounting Procedure
Install the each of the new part in the reverse order of removal procedure.
Figure 8-8 Removing the 17" LCD front cover
Table 8-2
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-1. Equipment
passed all required tests and is ready for use. “