GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
4 - 26
Section 4-7 - Keyboard Function Checks
8.) Check the 'Fast' radio button into Indicator box on scfpapitest program and press 'Set All' button.
R (Right)
Right Button
BodyPattern/Ellipse (Rotary
BodyPattern/Ellipse PushEncoder
Rotary
THI
Harmonics Button
Comment
comment Button
AO
Auto Right Button
Clear
Clear Button
B (Button)
B PushEncoder Button
TGC Slide
Slider : n=8 s1 ~ s8
B (Rotary)
B PushEncoder Rotary
NOTICE
When you can check the name displayed in log as you press or rotate keys, the main keyboard assy
was replaced correctly.
Figure 4-29 Fast button
NOTICE
After pressing Fast radio button, indicators should be blanked on the frontpanel.
Table 4-7
Button and Rotaty Table
Button or Rotary Name The Name displayed in Log
Button or Rotary Name
The Name displayed in Log