DIP Switch / Rotary Switch Setup
In this section, the settings of the DIP switch and rotary switch are explained.
DIP Switch Setup
The DIP switch is located under the maintenance cover at the side of the monitor.
SW No.
Default
Description (When ON)
1
OFF
Memory test mode at startup: ON
2
OFF
Not used. (Reserved function)
3
OFF
Not used. (Reserved function)
4
OFF
Not used. (Reserved function)
5 OFF
SpO
2
Body motion transmission via telemetry ON/OFF
6
OFF
NIBP test mode (Factory specification)
7
OFF
Not used. (Reserved function)
8 OFF
Enforced installing of application software and writing of
EEPROM data
1: This is the mode for factory inspection, and should not be used during normal operation.
5: Sets ON/OFF of SpO
2
body motion transmission during telemetry transmission.
Set according to the telemetry center capability on SpO
2
body motion information.
6: This is the mode for factory inspection, and should be set OFF during normal operation.
8: Used only for installing the software/EEPROM. Prevents unintended installing. Normally the
setting is OFF.
Rotary Switch Setup
The rotary switch is located under the maintenance cover at the side of the monitor.
No.
Default
Description
0
○
Normal operation mode
1
Not used.
2
Demo mode (Displays demo menu on the test menu.)
3
For Touch panel replacement.
4–5
Not used.
6
Analog adjustment when internal board is replaced.
(
Analog adjustment (user adjustment) menu is displayed in
the test menu.
)
7
Prohibited to use
8–D
Not used.
E
For EEPROM writing.
F
Initialize internal setup to factory default setup. (Initialize)
0: Normally, set the rotary switch to “0”.
2: To start the demo, press the demo start key on the test menu.
To enter the test menu, the password, “7000” is required.
3: Once the Touch panel has been replaced the default value (delta shift) needs to be
recalculated. Also if the calibration failed, the default value can be recalled.
6: The factory shipment adjustment value is preserved. Only adjust when internal board for
ECG, IBP, CO has been replaced.
7: Used for factory adjustment. If this mode is entered without adjustment information/facility,
and the equipment setup is changed, the equipment may become unusable. Do not use this
mode outside factory.
E: Enables to write equipment information to EEPROM.
F: When the power is turned ON with this mode, the user settings will be initialized.
7
-
2
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...