C A U T I O N
Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
Do not polish the housing with abrasive or chemical cleaner.
When sterilizing the entire room using a spray solution, pay close attention
not to have liquids get into the equipment or connectors.
Use only neutral detergent to clean the housing. Do not use chemical
cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent
and chemicals (cleanser, thinner, toluene, benzine, benzol, and synthetic
detergent for house and furniture), or sharp-edged tools. The surface resin
coating may be damaged, resulting in discoloration, scratches, and other
problems.
Do not reuse / re-sterilize the disposable airway adapter.
Do not open the housing.
Do not allow liquids such as alcohol or cleaning solution enter the
equipment or connectors.
If you accidentally wet the device, dry it completely and verify it operates
safely before usage.
Replace the components periodically as specified.
Precautions about the Wired Network System (DS-LAN
ΙΙ
)
W A R N I N G
Do not connect unspecified device to a wired network.
C A U T I O N
If performing wired network transmission, configure the display so that the
numeric data corresponded to the waveform is displayed. If not, the
displayed waveform or numeric data may not be transmitted.
The Operation Room ID is factory set to 000. If connected to the wired
network with the ID unchanged, monitoring on the central monitor will not
be possible.
When connecting to the wired network, verify that the Operation Room ID
does not duplicate with other monitors. Otherwise, monitoring on the
central monitor for both monitors will not be possible.
To connect to the DS-LAN
ΙΙ
network, set the Operating Room ID in the
range from 001 to 048.
There are following restrictions when connecting the DS-7000 to the
wired network.
Arrhythmia alarm of TACHY, BRADY, COUPLET, PAUSE,
TRIGEMINY will not be transmitted.
Arrhythmia alarm of “SLOW VT” will be transmitted as “VT”.
On a wired network, waveform, numeric data, alarm of TEMP3 will not
be transmitted. Also, the displayable waveform, numeric data, alarm
differs depending on the connected central monitor. Refer to the
operation manual for the respective central monitor.
If “HR/PR Alarm Source” is PR, and “PR Source” is BP, ECG waveform
will not be transmitted on a wired network. On the central monitor,
PR-IBP value will be displayed for HR. However, HR value from ECG
will be displayed for the NIBP list and ST measurement table.
If the “RR Alarm Source” selection is other than Impedance ,
respiration waveform will not be transmitted on a wired network.
If the “RR Alarm Source” selection is other than GAS , CO
2
waveform
will not be transmitted on a wired network.
For numeric data displayed as “
×××
”, maximum or minimum value of
measurable range will be transmitted.
The numeric data displayed as “
−
−
−
” will be treated as not measured
data.
xxii
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...