No
Check Item
Check Procedure
Criteria
12 Withstand
Voltage
Test (B-d)
Apply AC1500V for 1 minute between the
exterior and applied part (B-d).
Note: The voltage differs depending on
the internal protective circuit
composition of the equipment.
According to test procedure of IEC 60601-1 20.
Should withstand applied
voltage.