C A U T I O N
When DS-5800N/NX/NX
MB
is used as a central monitor, recall,
trend/table, data will not be displayed. Also,
Σ
recording cannot be
performed. For the ST display, overlap waveform will not be displayed on
the DS-5800N/NX/NX
MB
until 15 minutes elapses since the reference
waveform is set on the DS-7000.
As the DS-7000 do not have the arrhythmia template display and 12-lead
ST display function, these display on the central monitor will not be
corresponded.
If connected to a wired network, time/date will be the same with the
central monitor. Even if the time/date is changed on the DS-7000, it will be
corrected to the time/date of the central monitor.
If ECG lead (ECG1 or ECG 2) is changed on the DS-7000 while
monitoring ST display on the central monitor, the ST display will be
distorted. Redrawing the ST display will return the display to normal.
The respiration waveform and RR value based on the RR source selected
on the DS-7000 will be displayed on the central monitor. The monitoring
RR and APNEA will be the same as the one monitored on the DS-7000.
There are following restrictions when recording the DS-7000 data on the
central monitor recorder or the AU-5500N 8ch Recorder.
Only manual recording, alarm recording, periodic recording, and recall
recording can be performed on the AU-5500N.
When a parameter is not measured, the waveform for that parameter
will not be recorded, and measurement data will be recorded as “
−
−
−
”
or blank.
The measurement data displayed as “
×××
” will be recorded as “
−
−
−
”
on the central monitor recorder.
The QRS symbol, “S” printed on the built-in recorder will be printed as
“N” on the central recorder and AU-5500N.
For the waveform recording and trend recording, some parameters
may not be able to be recorded depending on the scale.
When performing table recording or trend recording on the central
recorder, some numeric data may not be recorded depending on the
parameter. Also, there are some trend scales that cannot be recorded.
If PR is set for the “HR Alarm Source” and BP is set for the “PR
Source”, ECG will not be recorded on the central recorder. PR-IBP
value will be printed instead for the HR value.
If the measurement unit of BP is kPa, the BP waveform, BP numeric
data, and NIBP numeric data will be treated as not measured data.
If the measurement unit of temperature is
°
F, the temperature data will
be treated as not measured data.
If the “RR Alarm Source” selection is other than Impedance ,
respiration waveform will not be output on the central recorder.
If the “RR Alarm Source” selection is other than GAS , CO
2
waveform
will not be output on the central recorder.
When trend recording, table recording, or NIBP list recording is output
on the central monitor recorder from the DS-7000, HR value derived
from ECG will be recorded for the HR value and ST trend.
xxiii
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...