The message, “Adjusting O
2
gain.” will be displayed.
To cancel the O
2
calibration, press the Cancel key.
C A U T I O N
If the O
2
offset has been adjusted, it is necessary to readjust the O
2
gain.
If the O
2
offset adjustment was not necessary, then the O
2
gain readjust
screen will not be displayed.
5
Update
the
O
2
calibration data.
When the O
2
gain adjustment is completed, the
message “Update calibration data?” will be displayed.
Stop supplying the calibration gas, and press the Yes
key.
Pressing the No key will cancel the O
2
calibration.
The message, “Updating calibration data.” will be
displayed.
During the updating process, O
2
calibration process
cannot be canceled.
8
-
5