Test Menu
In this section, functions of the test menu are explained.
To Display the Test Menu
Press the Menu
→
Maintenance
→
Test Menu keys and enter the password
(7000).
N O T E
The Test Menu screen may differ according to the software version.
Test Menu Items
Maintenance
: The hardware information of the equipment can be checked on this display.
Refer to the section on “Maintenance”.
Video I/F
: Displays various patterns of test screen to check the LCD operation. Refer
to the section on “Video I/F”.
Touch Panel
: Checks the touch panel key input operation.
Refer to the section on “Touch Panel”.
Touch Panel Adjustment :
Readjusts the touch panel if shifted.
Touch panel calibration will be performed.
Refer to the section on “Touch Panel Adjustment”.
Recorder Adjust
: When the recorder unit is replaced or repaired, readjustment will be
performed. Fine adjustment of the print density, paper out sensor, mark
sensor setup can be also performed.
The adjustment result will be stored. Refer to the section on “Recorder
Adjustment”.
QRS Detect
: If QRS detection is not correctly performed due to double trigger, etc.,
record Mode 1, Mode 2, Mode 3 waveforms and contact Fukuda Denshi.
Dyna Alert
: Analysis Information of Dyna Alert is displayed.
7
-
3
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...