2
Connect the specified calibration gas cylinder to the multigas unit.
When the preparation of the O
2
calibration is completed,
O
2
gain and O
2
offset value will be displayed, and the
message, “Start to supply calibration gas.” will be
displayed.
Supply the calibration gas, and press the OK key.
To cancel the O
2
calibration, press the Cancel key.
C A U T I O N
Use only the specified calibration gas. Proper calibration is not possible if
unapproved calibration gas is used.
The message, “Adjusting O
2
gain.” will be displayed.
To cancel the O
2
calibration, press the Cancel key.
C A U T I O N
If O
2
gain adjustment is started without supplying the calibration gas, the
message, “Check calibration gas.” will be displayed and O
2
gain adjustment
will cease.
8
-
3
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...