Periodic Check
●
Introduction
The periodic maintenance check is intended to check the medical equipment used daily in a
medical institution to prevent failures and accidents and to ensure safety and reliability.
As these are general check procedures covering the whole monitoring equipment, perform the
check procedure for the corresponded function of the equipment.
The check procedures are described for daily check and periodic check. Each check items must
be performed according to the described check procedure.
The consignee can select the check items according to the product quality, frequency of usage,
and maintenance check period. However, electrical safety items must also be performed.
For details of the electrical safety check procedure, refer to JIS T 0601-1, IEC 60601-1.
●
Service tools/Measurement Device
The measurement devices used for this periodic check are as follows.
Name Model
Type
Reference
Comprehensive Tester
SG-2000 or equivalent
Chunichi
ECG, RESP, BP, TEMP, CMRR
Waveform Simulator
217A or equivalent
DNI
NEVADA
ECG, RESP, BP, TEMP, CO
Leakage Measurement
Safety tester
3155/3156 or equivalent
HIOKI
Ground Leakage Current, Exterior Leakage Current
(Exterior - Line, Exterior - Ground, Exterior -
Exterior), Patient Leakage Current I/II/III, Patient
Measurement Current
Calibration Gas
80% O
2
(Po-03046630AIPD)
Fukuda
500cc Air Tank
NIBP
AC Resistance Tester
3157-01 or equivalent
HIOKI
Protective Ground Continuity Resistance
Telemetry Receiver
Receiver
Fukuda
Remote Control Unit
CF-700
Fukuda
Prepare the following devices as appropriate.
Name Model
Type
Reference
SpO
2
Simulator
Oxitest Plus or equivalent
An approximate check is possible by
measuring healthy subject.
DNI
NEVADA
SpO
2
NIBP Simulator
Cuff Link
An approximate check is possible by
comparing with auscultation on
healthy subject.
DNI
NEVADA
NIBP
Withstand Voltage Tester
3173 or equivalent
HIOKI
Withstand Voltage Test (A-a1, A-a2, B-a)
Reference Power Supply
Equipment
CA-100 or equivalent
Yokogawa
Oscilloscope
Multimeter
C A U T I O N
The measurement device must be properly calibrated.
7
-
35
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...