Cleaning the Housing
Clean the housing using tightly squeezed gauze or an absorbent cotton cloth dampened with
alcohol or a neutral cleanser.
C A U T I O N
Clean the equipment frequently so stains can be removed easily.
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
Do not allow liquids or cleaning solution to enter the monitor or connectors.
Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the case.
Do not polish the housing with abrasive or chemical cleaner.
When sterilizing the entire room using a spray solution, pay close attention
not to have liquids get into the monitor or connectors.
Use only neutral detergent to clean the housing. Do not use chemical cloth,
scrub brush, abrasive, polishing powder, hot water, volatile solvent and
chemicals (cleanser, thinner, benzine, benzol, and synthetic detergent for
house and furniture), or sharp-edged tools. The surface coating may be
damaged, resulting in discoloration, scratches, and other problems.
Cleaning the NIBP Cuff and Air Hose
Take out the bladder from the cuff bag, and wash the cloth bag with neutral detergent. After drying
it, put the bladder back in.
Disinfecting the Blood Pressure Transducers
Disinfect the blood pressure transducers according to the manufacturer’s guidelines.
Cleaning and Disinfecting the SpO
2
Sensor
【
NELLCOR
®
Sensor
】
Do not soak the sensor in water or antiseptic solution.
Wipe the DURASENSOR with disinfectant such as 70% alcohol. Do not disinfect by applying
radioactive rays, steam, or ethylene oxide.
OXISENSOR is a disposable sensor. Do not reuse or re-sterilize.
【
MASIMO
®
Sensor
】
Do not immerse the sensor or patient cable in water or cleaning solution. (Sensor and connector
are not waterproof.)
Do not disinfect the sensor and cable by irradiation, steam, or ethylene oxide.
Clean the Masimo reusable sensor (LNOP
®
DCI) and patient cable using the following procedure.
(1) Remove the sensor from the patient. Disconnect the patient cable from the sensor.
(2) Disconnect the sensor from DS-7000M.
(3) Wipe the sensor and cable using 70% isopropyl alcohol in cotton.
(4) Before reusing it, make sure it is dry completely to air.
The Masimo disposable sensor can be reused on a same patient if the light emitting and
receiving part is clean, and if it is still adhere to the skin.
The adhesiveness will return when the sensor is completely dried.
7
-
20
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...