Maintenance Check
Daily and Periodic Check
This section explains daily check and periodic check of the device.
About Maintenance Check
Periodic inspection must be performed. When reusing the device, which was left unused for a
while, always check that the device operates properly and safely before use.
To ensure safety, reliability, and high performance, a “Daily Check” and “Periodic Inspection” must
be performed. We are not liable for any accident arising from lack of maintenance.
C A U T I O N
Do not open the housing of this device.
Avoid alcohol or other liquids from getting into the device.
●
Daily Check
Perform daily inspection using the “Daily Check List” on the next page.
●
Periodic Check
The safety check conformed to IEC 60601 must be performed at least once every 2 years.
Periodic inspection of medical electronic equipment is mandatory to prevent failures and accidents
and to ensure safety and reliability.
Periodic maintenance may be performed by each medical institution or by a third party by
concluding a “Maintenance Contract”.
Periodic Replacement Period for Each Part
To ensure reliability, safety, function, and performance of this device, the following parts must be
replaced periodically.
Short Term Backup Battery
3 years
NIBP Unit
100,000 measurements
Recorder Unit
350 hours (Rec. accumulated time)
Gas Unit: Water Trap
6 months or when the water is half-full.
Gas Unit: CO
2
Absorber
1year
Gas Unit: O
2
Cell
1year
N O T E
The display panel utilizes exclusive fluorescent light for the backlight.
As this fluorescent light tube has product life cycle, it needs to be replaced
periodically.
C A U T I O N
The service life of the cooling fan used for this equipment is 40,000 hours.
However, if it is used with extremely high frequency in long duration, a
message such as “FAN Error” or “DS-7000 Hardware Error” may be
displayed or unusual sound may be generated from the cooling fan.
The periodic replacement parts must be replaced at specified period.
7
-
32
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...