The “TEMP not connected” message is displayed.
Cause
: When the temperature sensor is disconnected during temperature monitoring, this
message will be displayed.
Solution 1 : To cease monitoring, press the
(Alarm Silence) key to silence the alarm.
Solution 2 : To continue monitoring, plug in the temperature sensor. This will clear the message
and silence the alarm.
“TEMP Unit Error” message is displayed.
Cause
: A failure in the temperature unit is detected.
Solution
: Faulty unit can be considered.
Cardiac Output
The CO value varies more than
±
10% for consecutive measurements.
Cause 1 : The injection method is not appropriate.
Solution
: Inject within 1 to 3 seconds.
Cause 2 : Injection temperature is not appropriate.
Solution
: If iced injectate is used, pay attention not to warm the injector with your hands.
Cause 3 : The thermistor location is not appropriate.
Solution
: Reposition the thermistor by moving the catheter.
Cause 4 : Arrhythmia event has occurred during the measurement.
Solution
: Wait until the patient has a stable heart rhythm.
Cause 5 : There was patient body movement during measurement.
Solution
: Have the patient stay still during the measurement.
Cause 6 : The patient’s hemodynamics has changed during measurement.
Solution
: Wait until the patient has stable hemodynamics.
Abnormal measurement value is displayed.
Cause
: The catheter size, injectate volume, catheter constant (CC) is not correct.
Solution
: Set the proper condition, CC value for the used catheter.
The blood temperature (Tb), injectate temperature (Ti) is not displayed on the monitor.
Cause
: The catheter is not properly connected.
Solution
: Securely connect the catheter.
The thermodilution curve is deformed.
Cause
: The injection is not smooth, steady motion.
Solution
: Injection should be performed within 1 to 3 seconds.
6
-
13
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...