4. Re-insert the battery unit with the new battery.
Phillips screws M3 (
×
2)
・
Reconnect the connectors for the short/long term
backup battery.
5. Re-install the NIBP unit into the main body.
Phillips screws M3 (
×
4)
・
Reconnect the silicone tube, speaker cable and
NIBP cable.
6. Fill in the “Parts Replacement Label”.
Take out the “Parts Replacement Label” from the
“4/V250H Battery Kit”. Write down ”4/V250H Battery”
for “Parts Name”, and current date + 3 years for the
expiration date.
Attach the label on the patient monitor
7.
Press the Menu
→
Maintenance
→
Parts Usage Time keys and press the
Reset (password “7000”) key for “Short-Term Backup Battery”. Press the Yes (Hold 3sec.)
key and reset the usage months of short-term backup battery.
8. Press the following keys Menu
→
Maintenance
→
Test Menu
→
Maintenance to
confirm that the new battery is displayed and works properly.
7
-
26