The “Gas Unit Failure 4” message is displayed.
Cause
: The supply voltage to the gas unit is too low.
Solution
: Contact service department of Fukuda Denshi.
The “Sensor Cal. Error” message is displayed.
Cause
: The gas unit failed to calibrate the anesthetic agent detector.
Solution
: Contact service department of Fukuda Denshi.
The “Warming Up” message is displayed.
Cause
: The gas unit has not reached the full accuracy for anesthetic agent concentration.
Solution
: Wait until the warm up completes.
If the warm up does not complete within 30 minutes, check all the connected cables,
sampling tubes, nasal cannula, and turn OFF and ON the power again.
If the trouble persists, contact service department of Fukuda Denshi.
The “O
2
Sensor Failure” message is displayed.
Cause 1 : Faulty O
2
sensor inside the gas unit has been detected.
Solution
: Contact service department of Fukuda Denshi.
The “O
2
Module Failure” message is displayed.
Cause 1 : The gas unit selected on the serial communication setup menu does not correspond to
the connected gas unit.
Solution
: Select the correct gas unit on the serial communication setup menu.
Cause 2 : Faulty O
2
module inside the gas unit has been detected.
Solution
: Contact service department of Fukuda Denshi.
Ventilator
The “Vent. Data Unavailable” message is displayed.
Cause
: One of the following situations is observed.
・
Waiting to communicate with ventilator.
・
Waiting to communicate with ventilator after the power of the DS-7000 is turned ON.
・
”Vent. Data Unavailable” is displayed for the connection status message and
OFF is selected on the ventilator display.
Solution
: When normal communication is recovered, the communication status will be displayed
as “Vent. Online”.
The “Vent. Alarm” message is displayed.
Cause
: The following alarm has generated on the ventilator.
・
Parameter alarm such as AWP, MV, FiO
2
・
Technical alarm of the ventilator such as oxygen cell replacement
Solution
: Check the alarm cause on the ventilator, and take appropriate action.
6
-
16
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...