2
-
7
Sevoflurane 0.3 vol. %
Desflurane 0.3 vol. %
Mixed Gas Threshold : 0.2 vol. % + 10 % of total concentration
Rise Time
: CO
2
: 350msec
N
2
O: 400msec
O
2
:
600msec
Agents: 450msec
Occlusion Clearing
: Automatic
Auto Zeroing
: Occurs 30 to 60 minutes
Duration: 3.0 to 7.0 seconds
Manual user calibration not required
Warm-up Time
: 1 minute to first waveforms
<20 minutes to full accuracy
Operating Temperature : 15 to 35
°
C
Operating Humidity
: 30 to 85% (non-condensing)
Storage Temperature : –5 to 50
°
C
Storage Humidity
: 15% to 90% non-condensing
Altitude
: –1,000 to 10,000ft (–300 to 3000m)
Interfering Gases for Anesthetic Agents
The monitor will report small changes in agent concentrations when anesthetic agents and other
gases are used.
Expected agent changes are provided here for the purpose of comparison.
For Gas Mixtures of 5% CO
2
Agents
Agent Volume
Change of CO
2
N
2
O 60%
+0.2
Halothane 4%
+0.1
Enflurane 5%
+0.1
Isoflurane 5%
+0.1
Sevoflurane 5% +0.2
Desflurane* 5% 0.0
Helium 50%
0.0
Ethanol 1%
+0.1
Isopropanol 1% 0.0
Acetone 1%
+0.1
Methane 1%
0.0
Metered dose inhaler propellant
1%
0.0
80% Xenon
Not intended for use with Xenon
N/A
For Gas Mixtures of 30% O2
Agents
Agent Volume
Change of CO
2
N
2
O 60%
–0.1
Halothane 4%
–0.2
Enflurane 5%
–0.2
Isoflurane 5%
–0.2
Sevoflurane 5% –0.2
Desflurane* 5% –0.2
Helium 50%
–0.3
Ethanol 1%
–0.2
Isopropanol 1% +0.5
Acetone 1%
+0.1
Methane 1%
–0.1
Metered dose inhaler propellant
1%
+0.2
80% Xenon
Not intended for use with Xenon
N/A
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...