Disassembly and Assembly
1. Remove the NIBP unit (bottom) from the main
body.
Phillips screws M3 (x4)
・
Take care when disconnecting the silicon tube, the
speaker cable, and the NIBP cable.
2. Remove the battery unit.
Phillips screws M3 (
×
2)
・
Disconnect the short/long time backup
connectors.
3. Assemble the "4/V250H battery exchange kit".
Phillips screws M3 (
×
1)
・
The short-term backup battery is stored in the
battery cover.
・
The cover is fixed with the bracket and the screw.
Kit content:
1) Short-term backup battery
(250H-073676(NI-MH)) P/N: 1U1019A
2) Parts Replacement Label
3) Tie wrap (
×
2)
・・・
not used.
4) Battery stand (TMS-20)
P/N: 6C0077470
・・・
not used.
5) Battery cover (blue)
P/N: 6C0077480
6) Instruction manual
7
-