CF Card
By using a CF card, the data on this equipment can be stored, read, or copied.
CF Card Format
Formats the CF card so that it can be used on the DS-7000 series.
C A U T I O N
A CF card formatted on a PC cannot be used on the DS-7000.
The CF card must be formatted on the DS-7000.
DS-7000
→
CF Card
Stores the data to the CF card.
All Data
: Stores all data.
Setup Data
: Stores the following data.
Alarm Mode
: Stores the alarm data.
Display Mode
: Stores the display data.
Remote Control
: Stores the remote control data.
Patient Data
: Stores the patient information data.
LOG Data
: Stores LOG data for error analysis.
CF Card
→
DS-7000
Reads the data from the CF card.
All Data
: Reads all data.
Setup Data
: Reads the following data.
Alarm Mode
: Reads the alarm data.
Display Mode
: Reads the display data.
Remote Control
: Reads the remote control data.
Patient Data
: Reads the patient information data.
C A U T I O N
The hospital name, operation room ID, etc. are also copied to the CF card
from this menu. Use this function before software upgrade and repairs.
To transfer data between monitors (DS-7000), go to Menu then select
CF Card .
7
-
14
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...