2
-
10
Other Interference
Ether, cyclopropane and methoxyflurane are contraindicated for use with the multigas unit. The
effect of ethyl alcohol and metabolic ketones and acetone are negligible.
W A R N I N G
The multigas unit is not intended for monitoring gas mixtures containing
methoxyflurane or halogenated hydrocarbons not specifically listed as a
monitored gas.
Cardiac Output
Measurement Method : Thermodilution Method
Measurement Range : 0.1–20L/min
Measurement Accuracy : 0.1 to 10L/min:
less than
±
10%
10 to 20L/min:
less than
±
15%
Measurement Temperature Range and Accuracy
Blood Temperature
: 17–45
°
C
±
0.3
°
C
Injectate Temperature : –1–35
°
C
±
0.5
°
C
Recording
Recording Speed
: 50mm/s, 25mm/s (Error: within
±
5%)
Resolution
Head Direction : 8 dots/mm
Feed Direction : 40 lines/mm (at recording speed of 25mm/s)
Rec. Waveform
: 3 waveforms
Rec. Type
: Waveform Recording, Table Recording, Graphic Recording
Detection
: Paper out, page mark, paper cassette error, printhead temperature
Protective Circuit
: Printhead over current, printhead overheating, motor over current,
surge current
Analog Waveform Output
Output Voltage
: ECG output 1V/mV (fixed), BP output 1V/100mmHg (fixed)
Output Voltage Accuracy : within
±
10% (Both ECG and BP output)
Delay Time
: 35ms and below (Both ECG and BP waveform)
Frequency Range
: ECG output 0.05Hz–30Hz
BP output DC–30Hz
Pacemaker Pulse
: No Pacemaker Pulse
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...