Calibration of Multigas Unit (MGU-701)
The O
2
calibration is required every time the O
2
cell (MGU-701) is replaced. (Refer to “9.
Maintenance”).
To Display the GAS Calibration Screen
Press the
Menu
→
Maintenance
→
GAS Calibration keys.
Gas Calibration Procedure
1
Press the Start O
2
Cal key.
The message, “Preparing for O
2
calibration.” will be
displayed.
Once the preparation has started, the O
2
calibration
process cannot be cancelled.
N O T E
The O
2
calibration cannot be started until the multigas unit has finished its
“warming up” stage.