Replacing the Recorder Unit
●
Service Life of the Recorder Unit
The recorder unit can be used for approximately 350 hours.
If using the recorder over its service life, the print out will definitely become worse and the recorder
will need to be replaced.
●
Tools for Replacement
・
Phillips Screwdriver (Large)
●
Disassembly and Assembly
Replace the unit with the following procedure by paying attention to the connectors and tubing.
1. Remove the air filter and then the rear panel
(6 screws).
2. Remove the recorder unit (3 screws) and
replace it with a new one (inverse procedure).
3. Re-install the rear panel and air filter.
4. Press the Menu
→
Maintenance
→
Parts Usage Time keys and press the
Reset (password “7000”) key for “Recorder
Operation Time”. Press the Yes (Hold 3sec.)
key to reset the usage for the Recorder
Operation Time.
5. Turn the power ON and check that the recorder operates properly.
7
-
28
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...