Sensor” message is displayed.
Cause
: The sensor is detached from the patient.
Solution 1 : Check if the sensor is properly attached to the patient.
Solution 2 : Check if the light emitting part and light receiving part of the sensor LED are aligned.
The “Pulse Search” message is displayed.
Cause
: The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution
: Check if the light emitting part and light receiving part of the sensor LED are aligned.
The “No Pulse Detect” message is displayed.
Cause
: The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution
: Check if the light emitting part and light receiving part of the sensor LED are aligned.
The “Motion Artifact” message is displayed.
Cause
: There is excessive body motion from the patient.
Solution
: Change the sensor position where the body motion will have less effect.
The pulse waveform is not displayed, or interrupted.
Situation : The “Check SpO
2
sensor” is displayed.
Cause 1 : The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution
: Check if the light emitting part and light receiving part of the sensor LED are aligned.
Cause 2 : Sensor is defective.
Solution
: Replace the sensor.
Cause 3 : SpO
2
sensor is not firmly connected to the SpO
2
input connector.
Solution
: Make sure the SpO
2
sensor is securely connected.
Cause 4 : Sensor is exposed to light.
Solution
: Place a black or dark cloth over the sensor to avoid direct sunlight. Also when not used,
avoid placing the sensor in light or unplug the sensor from the connector.
The SpO
2
measurement is unstable.
Cause
: There is excessive body motion from the patient, which disables correct
measurement.
Solution
: 1. Have the patient lie still as much as possible.
2. Relocate the sensor, or change the sensor to which the body motion will have less
influence.
The “SpO
2
Unit Error” message is displayed.
Cause 1 : There is a problem in the communication with the SpO
2
measurement unit.
Solution
: A defective cable or faulty SpO
2
unit can be considered.
Reference : When “SpO
2
unit error” is displayed on the home display, the detail of the error will be
displayed at the RECOVERY NUMBER of SpO
2
in the test menu.
“SERIOUS ERROR”
“MODULE ERROR REQUIRES RESET”
“TRANSIENT ERROR”
“COMMUNICATION ERROR”
6
-
10
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...