Module Installation
This is where you install the telemeter (HLX-501, HLX-561) software for the DS-7000.
1. Insert the CF card of the software upgrade of the module.
2. The current software version and the CF card version are displayed, then press the
Start Install key to install the software.
C A U T I O N
During the install process, do not turn off the power of the equipment.
●
RTC Setup
Sets the built-in RTC (clock).
1. Use the numeric keypad to enter the date and time.
2. Press the Yr , Mo , Day , Hr , Min key, and set each item.
3. Press the Setup key to store the setup.
C A U T I O N
This RTC Setup in the test menu is the reference time setting at factory
shipment. Changing the RTC setting will affect the accumulative hours
calculation, etc. Therefore, do not change this setting other than when
replacing the lithium battery or a board.
For normal time setting, go to Menu
→
Settings
→
Time/Date .