The “SpO
2
Sensor Defective” message is displayed.
Cause
: Sensor is defective.
Solution
: Replace the sensor.
The “Disconnect” message is displayed.
Cause
: When the SpO
2
relay cable is disconnected during SpO
2
monitoring, this message will
be displayed.
Solution 1 : To cease monitoring, press the
(Alarm Silence) key and silence the alarm.
Solution 2 : To continue monitoring, plug in the SpO
2
relay cable. This will clear the message and
silence the alarm.
SpO
2
(Masimo
®
Model; DS-7000M)
The “SpO
2
Sensor Defective” message is displayed.
Cause
: Sensor is defective.
Solution
: Replace the sensor.
The “Check SpO
2
Sensor” message is displayed.
Cause
: The sensor is detached from the patient.
Solution 1 : Check if the sensor is properly attached to the patient.
Solution 2 : Check if the light emitting part and light receiving part of the sensor LED are aligned.
The “SpO
2
Low Perfusion” message is displayed.
Cause
: The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution
: Check if the light emitting part and light receiving part of the sensor LED are aligned.
The “Pulse Search” message is displayed.
Cause
: The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution
: Check if the light emitting part and light receiving part of the sensor LED are aligned.
The “Check SpO
2
Sensor” message is displayed.
Cause
: The sensor is exposed to too much ambient light. The detecting part of the sensor is
not covered appropriately.
Solution 1 : Turn down or turn off the light.
Solution 2 : Avoid sensor from exposure to ambient light.
Solution 3 : Relocate the sensor position.
The “SpO
2
Noise” message is displayed.
Cause
: External signal or energy is interfering with the measurement.
Solution
: Remove the external interference.
The “Unknown SpO
2
sensor” message is displayed.
Cause
: Unrecognizable sensor is connected.
Solution
: Replace the sensor with a specific one.
The “SpO
2
Low Signal IQ” message is displayed.
Cause
: There is excessive body motion or the position of the sensor is not appropriate.
Solution 1 : Check if the light emitting part and light receiving part of the sensor LED are aligned.
Solution 2 : Change the sensor position where the body motion will have less effect.
6
-
11
Содержание DynaScope 7000 Series
Страница 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Страница 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 70: ...4 4 Blank Page ...
Страница 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Страница 92: ...6 18 Blank Page ...
Страница 143: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100110 200805 ...