Operator’s Manual: ThermoJet
Chapter 3: System Operation
© SP Industries, Inc.
200-31974-999 Rev. 009
17
pan is recommended. Should fluid accumulate, this is a sure sign of poor supply air quality.
Supply air with a low dew point and low oil carry-over, will cause little or no liquid
accumulation. Conversely, rapid or consistent liquid accumulation indicates poor air supply
performance. Continued use with this condition will cause permanent damage to the
Desiccant Dryer (internally mounted) and premature equipment failure.
The drain line exiting the machine may be re-routed to an alternate drain site at the operator’s
discretion. If this option is chosen, the operator should make provisions to assess the air
supply quality on a regular schedule.
RS232 Communication
To communicate using the RS232 protocol, unplug the converter cable (c) and connect your
communication cable directly to the RS232 port.
GPIB/IEEE488 Communication
To communicate using the GPIB protocol, the converter cable (c) must be connected to the
RS232 port. Then the operator must connect the in-house GPIB cable into the ThermoJet
IEEE/488 port.