Operator’s Manual: ThermoJet
Chapter 5: Display Screens
© SP Industries, Inc.
200-31974-999 Rev. 009
50
I. Temperature Calibration Screen 2 –
This calibration procedure will calibrate the nozzle only.
This procedure is not intended for and will not calibrate the DUT. From the Utilities Screen (Fig.
16), selecting the “Temperature Calibration 2” button displays the Start screen below. Before
calibration, the system should be in Nozzle control mode, and be controlling in Idle mode.
Calibration procedure details are presented in more detail in the calibration section.
Figure 37 – Air Temperature Calibration Screen 2
Operator specified
setpoint
Back to Manual Mode
screen
Back to Utilities screen
Operator specified
setpoint
Temperature Calibration 2 Start Screen