Operator’s Manual: ThermoJet
Chapter 5: Display Screens
© SP Industries, Inc.
200-31974-999 Rev. 009
32
Figure 16 – Utilities Screen
A.
Auto-Tuning Screen -
From the Utilities screen (Fig. 16), select the “Auto-Tuning” button and
the screen depicted below will open. From Auto-Tune the operator can:
•
Set the auto-tune temperature values
•
Set the air flow rate
NOTE:
The maximum allowable air temperature must always be greater than the desired
setpoint, and the minimum allowable air temperature must always be less than the desired
setpoint. In general, the wider the allowable temperature range, the better the resulting loop
performance and temperature transition rate of the device under test (DUT).
Figure 17 – Auto-Tune Screen
Utilities Screen
Status window displays
the system’s current state
Main menu buttons
Auto-Tuning Screen
(DUT only
)
Set temperature
setpoint around which
auto tuning will be
Set minimum allowable
air temperature,
default = -85°C
Access Air Flow Screen
Main menu buttons
Set maximum allowable
air temperature,
default = -225°C
Start/Stop auto-tune
sequence
PID values from auto
tuning