6270A
Service Manual
84
Module Calibration
For the procedures in the subsequent sections, environmental conditions must
be within the correct limits for the type of laboratory. Make sure to exercise the
UUT range before calibration and do As Found measurements before C0 or C1
adjustments, repairs, or upgrades are made to an operational instrument.
Note that each slot in the module bay (left to right, slot 1 on left facing front
panel) has a specific number value assigned to it:
•
Slot 1 = 5
•
Slot
2=15
•
Slot
3=25
•
Slot
4=35
•
Slot
5=45
Calibration of Modules in Chassis
To calibrate the modules in the chassis:
1. Zero the module.
2. Record the current C0, C1, zOffset levels that you see on the front panel
(values are rounded up. The resolution provided by the remote
communication method may be necessary).
a. Select
SETUP>Module Information>Slot
(n) - where (n) is the slot
number
b. Record:
•
Calibration
Date
•
C0
•
C1
•
zOffset.
3. Record the current C0, C1, zOffset levels from the remote communication
via RS232 (null modem) or IEEE.
Below is an example of reading the C0 command from pressure module in slot 1:
Query Example:
CAL:PRES5:DATA:VAL1?
Response:
C0,-5.16999997E-02
Query Example:
CAL:PRES5:DATA:VAL2?
Response:
C1,1.0000
Query Example:
CAL:PRES5:ZERO:VAL?
Response:
+2.13499999E-03
a. Read C0 of slot
•
CAL:PRES(n):DATA:VAL1? - where (n) is the correct value for slot to
be read.
b. Read C1 of slot
•
CAL:PRES(n):DATA:VAL2?
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...