Pressure
Controller/Calibrator
Troubleshooting
115
Control
The Control diagnostics menu shows information about the PCM:
•
Port Pressure
–
This menu shows the pressure measurement at the test,
supply, and exhaust ports.
•
Control Module
– This menu shows the PCM model number, serial number,
and firmware version.
•
Control Coefficients
– This menu shows the control coefficients (C0) the
PCM uses. To change the coefficients, tune the PCM with the autotune
function located in this menu.
•
Autotune
–
Autotune is a fully-automated procedure that runs the Product
through a series of pressures across its ranges. The Product then alters the
control coefficients to provide the best control performance. Only use
autotune when the control performance is unacceptable. The system must be
leak free, have minimal external volume, and the sealing plug on the test port
must be rated for the maximum pressure of the system.
1. Plug the test port and make sure connections are capable of accepting
pressures of 90 % FS of the highest module installed in chassis.
2. Select
SETUP>Tasks>Leak Test
, refer to
Leak Test
.
Note
With the test port plugged and no external volume, System Leak is
expected to achieve better than 0.04 %FS/Min rate.
3. Select
SETUP>Diagnostic>Control>Primary>Autotune
.
This process should take 20 minutes to 40 minutes. If there are leaks or the
volume is too large to run autotune, then system will error or timeout.
Leak Test
To perform a basic leak test, select the pressure range then select
SETUP>Tasks>LeakTest
. Complete the following entries:
a. Setpoint – Pressure set point at which to perform the leak test
b. Stability – After achieving the setpoint, the time in seconds to wait for stability
before the Duration of the test begins.
c. Duration – The time in seconds to collect data for the leak-rate calculation
d. Mode – Defaults to external, not selectable
e. Run – Starts the test
f. Exit – Exits the leak test
Note
With the test port plugged and minimal external volume, system leak
is expected to achieve less than 0.04 % FS/minute change. To
resolve this performance, stability time must be set to allow thermal
changes to stabilize. This time is typically greater than 60 seconds
and up to 240 seconds, depending on testing conditions.
To assist with evaluation of internal leak and valve performance, use the
Pneumatic Self-Test function located on the Measure Diagnostic screen.
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...