78
Calibration and Adjustment Procedures
This procedure describes calibration and performance verification to the
specifications, As Found data collection procedures, adjustment, and as left
verification procedures for the Product. The procedure is meant to bring the
Product into specification after repair or when the UUT fails the performance
tests.
•
Do the As Found data collection with the described setups and pressure set
points. The Product is left in measurement mode and pressure is applied with
the appropriate reference.
•
The adjustment procedure can be done in chassis or with a calibration sled,
or COMPASS for Pressure software.
•
Do the as left (verification) like the As Found data collection.
Required Equipment
The minimum use specifications in Table 6 are the principal parameters required
for performance of the calibration. Use the minimum specifications to assist in the
selection of alternate equipment, which may be used at the discretion of the
laboratory.
Table 6. Required Equipment
Item
Minimum Use
Specifications
Equipment Notes
Force Balance
Piston Gauge
(FPG)
Range: -15 KPa to
15 KPa
FPG8601
For Bi-Directional PMMs
≤
15 kPa FS
Piston Gauge
System
Range: As required by
P-C selection
PG7000 or 2465
N/A
Piston-Cylinder
PM200: Table 4
PM600: Table 5
PM200: Table 4
PM600: Table 5
N/A
Mass Set
Range: As required by
P-C selection
AMH or Manual
N/A
Mass Carrying Bell
Range: As required by
P-C selection
PG7000 or 2465
N/A
Piston Gauge
Monitor
Range: As required by
P-C selection
2455 or 2456 PGM
2465 system only
Environment
Monitor
Range: As required by
P-C selection
2456-LEM
2465 system only
Temperature
Probe
Range: As required by
P-C selection
PG7000 or 2465
N/A
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...