Pressure
Controller/Calibrator
Calibration and Adjustment Procedures
79
Item
Minimum Use
Specifications
Equipment Notes
Barometric
Reference
Range: 70 KPa to
110 kPa
Barometric range RPM3
or RPM4
Must interface with Piston
Gauge or COMPASS for
Pressure
Instrument
manuals, work
instructions,
specifications, and
other
documentation for
any associated
equipment.
Software -
Required if you are
not using the on-
board calibration
routine in the
Product menu.
•
COMPASS for
pressure calibration
software
•
COMPASS for
pressure unit of
measure converter
•
(version 5.x.0 or
newer)
•
(version x.0x Build x or
newer).
Standard PC
Windows 7, 8, or 10
Level vial
To estimate verticality to
the acceleration of gravity.
Ruler
To measure head height
differences between the
reference and the UUT.
Pressure tubing
and fittings as
required and
appropriate for the
pressure range
and system.
Calibration procedures described in this section, will need:
•
If a PG7000 Piston Gauge is used:
o
Calibrated and approved piston gauge (PG) base platform.
o
Calibrated and approved piston-cylinder for the range to be calibrated.
o
Calibrated and approved mass set compatible with P-C and PG base
being used and capable of generating the force necessary for the
recommended calibration points.
•
If a FPG8601 is used:
o
Calibrated and approved FPG.
o
Test pressure controller.
o
Dual volume unit (DVU).
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...