6270A
Service Manual
76
Calibration Adjustment
PM200
PM200
Range
MIN
Pressure
MAX
Pressure
Reference
System
1
Piston Cylinder
1
Reference
Mode
Dwell
Time
Averaging
Time
----------- [kPa] [kPa] -------------
-------------
-------------
[s]
[s]
G20M
0
20 000
PG7202
200 kPa, 500 kPa, 1 MPa, 2 MPa /kg P-C Gauge
60
30
G14M
0
14 000
PG7202
200 kPa, 500 kPa, 1 MPa, 2 MPa /kg P-C Gauge
60
30
G10M 0
10
000 PG7102 100
kPa/kg
P-C
Gauge
60 30
G7M 0
7000 PG7102 100
kPa/kg
P-C
Gauge
60 30
G4M 0
4000 PG7102 100
kPa/kg
P-C
Gauge
60 30
G3.5M 0
3500
PG7601 100
kPa/kg
P-C
Gauge
60
30
G2.5M 0
2500
PG7601 100
kPa/kg
P-C
Gauge
60
30
G2M 0
2000 PG7601 100
kPa/kg
P-C
Gauge
60 30
G1.4M 0
1400
PG7601 50
kPa/kg
P-C
Gauge
60
30
G1M 0
1000 PG7601 50
kPa/kg
P-C
Gauge
60 30
G700K 0
700
PG7601 20
kPa/kg
P-C
Gauge
60
30
G400K 0
400
PG7601 20
kPa/kg
P-C
Gauge
60
30
BG250K 0
250
PG7601
10
kPa/kg
P-C
Gauge
60
30
BG250K
2
-100
0
PG7607,
2465
5 kPa/kg P-C, (TL) P-C
Differential, Negative
Gauge
60 30
BG200K 0
200
PG7601
10
kPa/kg
P-C
Gauge
60
30
BG200K
2
-100
0
PG7607,
2465
5 kPa/kg P-C, (TL) P-C
Differential, Negative
Gauge
60 30
A200K 2
200
2465
(TL)
P-C
Absolute
by
Vacuum 60
30
BG100K 0
100
PG7607,
PG7601,
2465
5 kPa/kg P-C,
10 kPa/kg P-C, (TL) P-C
Gauge 60
30
BG100K
2
-100
0
PG7607,
2465
5 kPa/kg P-C, (TL) P-C
Differential, Negative
Gauge
60 30
A100K 2
100
2465
(TL)
P-C
Absolute
by
Vacuum 60
30
BG60K -60
60
PG7607,
PG7601
5 kPa/kg P-C,
10 kPa/kg P-C
Differential 60
30
BG40K -40
40
PG7607,
PG7601
5 kPa/kg P-C,
10 kPa/kg P-C
Differential 60
30
BG35K -35
35
PG7607,
PG7601
5 kPa/kg P-C,
10 kPa/kg P-C
Differential 60
30
BG2.5K -2.5
2.5
FPG8601 FPG8601
Gauge
90
30
1.
A reference may be substituted if the recommended reference is unavailable. The substituted reference must be able to cover the MIN and MAX
pressures required for the calibration of the applicable PMM.
The references given here are the preferred from an overall uncertainty and calibration
interval viewpoint. Other types of references, such as a Fluke Calibration 7250 Pressure Controller/Calibrator, can potentially be used when the
standard is maintained on a shorter calibration interval.
2.
Negative Gauge Calibration. Use PG Differential Mode of reference is PG7601/PG7607. Use 2465 Negative Gauge Mode if reference is 2465. -
100 kPa is the nominal minimum pressure. Depending on ambient laboratory pressure, the laboratory may not be able to achieve -100 kPa.
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...