6270A
Service Manual
8
PM600 Modules
The product specifications describe the Absolute Instrumental Uncertainty of the Product. The product specifications
include linearity, hysteresis, repeatability, resolution, reference standard measurement uncertainty, 1 year stability, and
temperature effects from 15 to 35 °C. The product specifications are provided at a 95 %, k=2, normally distributed, level of
confidence.
Model
Gauge Mode
Range
(SI Units)
Absolute Mode
Range
(SI Units)
Gauge Mode
Range
(Imperial Units)
Absolute Mode
Range
(Imperial Units)
1-Year Specification
Relative
Specification
(% Reading)
Threshold
Specification
(% Span)
Absolute Mode
Adder
(% Full Scale)
PM600-
BG15K
-15 kPa to
15 kPa
-
-60 inH
₂
0 to
60 inH
₂
0
-
0.01 %
0.003 %
-
PM600-
G100K
0 kPa to
100 kPa
-
0 psi to 15 psi
-
0.01 %
0.003 %
-
PM600-
G200K
0 kPa to
200 kPa
-
0 psi to 30 psi
-
0.01 %
0.003 %
-
PM600-
A100K
-100 kPa to
0 kPa
6 kPa to 100 kPa
-13.8 psi to
0 psi
0.9 psi to 15 psi
0.01 %
0.003 %
0.007 %
PM600-
A200K
-90 kPa to
100 kPa
10 kPa to 200 kPa
-13.2 psi to
15 psi
1.5 psi to 30 psi
0.01 %
0.003 %
0.007 %
PM600-
A350K
-90 kPa to
250 kPa
10 kPa to 350 kPa
-13.2 psi to
35 psi
1.5 psi to 50 psi
0.01 %
0.003 %
0.007 %
PM600-
A700K
-82 kPa to
700 kPa
18 kPa to 700 kPa
-12.1 psi to
100 psi
2.6 psi to
100 psi
0.01 %
0.003 %
0.007 %
PM600-
A1.4M
-0.065 MPa to
1.4 MPa
0.035 MPa to
1.4 MPa
-10 psi to
200 psi
5 psi to 200 psi
0.01 %
0.003 %
0.007 %
PM600-
A2M
-0.03 MPa to
2 MPa
0.07 MPa to
2 MPa
-5 psi to
300 psi
10 psi to 300 psi
0.01 %
0.003 %
0.007 %
PM600-
A3.5M
-0.03 MPa to
3.5 MPa
0.07 MPa to
3.5 MPa
-5 psi to
500 psi
10 psi to 500 psi
0.01 %
0.003 %
0.007 %
PM600-
A7M
0 MPa to 7 MPa
atmosphere to
7 MPa
0 psi to
1000 psi
atmosphere to
1000 psi
0.01 %
0.003 %
0.007 %
PM600-
A10M
0 MPa to
10 MPa
atmosphere to
10 MPa
0 psi to 1500
psi
atmosphere to
1500 psi
0.01 %
0.003 %
0.007 %
PM600-
A14M
0 MPa to
14 MPa
atmosphere to
14 MPa
0 psi to
2000 psi
atmosphere to
2000 psi
0.01 %
0.003 %
0.007 %
PM600-
A20M
0 MPa to
20 MPa
atmosphere to
20 MPa
0 psi to
3000 psi
atmosphere to
3000 psi
0.01 %
0.003 %
0.007 %
Notes
•
Gauge mode uncertainty is the greater of the relative uncertainty and the threshold uncertainty. For absolute ranges used in gauge
mode there is an additional uncertainty of ±7 Pa for dynamic barometric compensation. When combined with other uncertainties this
changes the threshold uncertainty for the PM600-A100K to 0.008 % Span and for the PM600-A200K to 0.004 % Span.
•
For absolute mode the % span for Threshold Uncertainty the span is calculated from zero to the full scale of that range.
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...