6270A
Service Manual
74
7
8
9
+/-
4
5
6
ESC
1
2
3
ENTER
0
.
SET P
VENT
PUSH
TO ENTER
SELECT
W
REMOTE
PPC4
PRESSURE CONTROLLER/CALIBRATOR
A700K/A160K
Test
Reference
Supply
Measured Pressure (Test Port)
Supply Pressure Port - 115% FS or
FS+15 psi whichever is greater
hxn310.eps
Figure 44. PM600 Barometer Optimization Setup
The UUT is set up with a minimum pressure of 75 kPa and maximum of 103 kPa.
Use the serial number of the PM600 as the UUT Serial Number. The test
definition is shown in Table 8.
Table 5. PM600 internal On-board Barometer Pressure Point Matrix
As Found Points / As Left Points
Calibration
Segment
Pressure
No.
Set Pressure
[kPa]
Dwell Time
[s]
Averaging Time
[s]
Ascending Points
1 75
60
30
2 103
60
30
To clear the current C0 and C1:
a. Set C0 to zero with the command:
OFFSET_SET2 0
b. Set C1 to one with the command:
GAIN_SET2 1
c. Set the calibration date with the command: CAL_DATE2 yyyymmdd
d. Save with the command:
CAL_STORE
Note
You must send the Store command.
For a NEW C0 and C1:
Run two points of 75 kPa and 103 kPa and calculate the new C0 and C1 and
send to the module and save the new values.
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...