Pressure
Controller/Calibrator
Control Specifications
7
PM200 Modules
Model
Range (SI Units)
Range (Imperial Units)
Measurement Mode
1-Year
Specification
(%FS)
PM200-BG2.5K
-2.5 kPa to 2.5 kPa
-10 inH
₂
0 to 10 inH
₂
0
gauge 0.20
%
PM200-BG35K
-35 kPa to 35 kPa
-5 psi to 5 psi
gauge
0.05 %
PM200-BG40K
-40 kPa to 40 kPa
-6 psi to 6 psi
gauge
0.05 %
PM200-A100K
2 kPa to 100 kPa
0.3 psi to 15 psi
absolute
0.10 %
PM200-BG100K
-100 kPa to 100 kPa
-15 psi to 15 psi
gauge
0.02 %
PM200-A200K
2 kPa to 200 kPa
0.3 psi to 30 psi
absolute
0.10 %
PM200-BG200K
-100 kPa to 200 kPa
-15 psi to 30 psi
gauge
0.02 %
PM200-BG250K
-100 kPa to 250 kPa
-15 psi to 36 psi
gauge
0.02 %
PM200-G400K
0 kPa to 400 kPa
0 psi to 60 psi
gauge
0.02 %
PM200-G700K
0 kPa to 700 kPa
0 psi to 100 psi
gauge
0.02 %
PM200-G1M
0 MPa to 1 MPa
0 psi to 150 psi
gauge
0.02 %
PM200-G1.4M
0 MPa to 1.4 MPa
0 psi to 200 psi
gauge
0.02 %
PM200-G2M
0 MPa to 2 MPa
0 psi to 300 psi
gauge
0.02 %
PM200-G2.5M
0 MPa to 2.5 MPa
0 psi to 360 psi
gauge
0.02 %
PM200-G3.5M
0 MPa to 3.5 MPa
0 psi to 500 psi
gauge
0.02 %
PM200-G4M
0 MPa to 4 MPa
0 psi to 580 psi
gauge
0.02 %
PM200-G7M
0 MPa to 7 MPa
0 psi to 1000 psi
gauge
0.02 %
PM200-G10M
0 MPa to 10 MPa
0 psi to 1500 psi
gauge
0.02 %
PM200-G14M
0 MPa to 14 MPa
0 psi to 2000 psi
gauge
0.02 %
PM200-G20M
0 MPa to 20 MPa
Psi 0 to 3000 psi
gauge
0.02 %
Notes
•
Gauge mode modules (PM200-GXXX or PM200-BGXXX) with ranges of 100 kPa (15 psi) or greater will support absolute mode
measurement when used with a Barometric Reference Module.
•
For temperatures from 15 °C to 18 °C and 28 °C to 35 °C, add 0.003 % FS/°C. Uncertainty for gauge mode modules assumes
routine zeroing. Uncertainty for absolute mode modules includes 1-year zero stability. This specification can be reduced to
0.05 % FS if the PM200 module is zeroed on a continuing basis to remove the 1-year zero stability component. Instrumental
Measurement Uncertainty for gauge mode modules used in absolute mode by addition of a barometric reference module is
calculated as the uncertainty of the gauge mode module plus the uncertainty of the barometric reference module.
•
Instrumental Measurement Uncertainty for gauge mode modules used in absolute mode by addition of a barometric reference
module is calculated as the uncertainty of the gauge mode module plus the uncertainty of the barometric reference module.
•
For temperatures from 15 to 18 C and 28 to 35 C, add 0.003 % FS/C.
Содержание 6270A
Страница 5: ...Contents continued iii Inspect and Replace the CPS valves 122 Replacement Parts 124 Screw Torque 136...
Страница 6: ...6270A Service Manual iv...
Страница 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Страница 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Страница 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Страница 37: ...Pressure Controller Calibrator Module Guide Rails 31 1 2 3 hxn209 eps Figure 13 Remove Module Guide Rails...
Страница 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Страница 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Страница 47: ...Pressure Controller Calibrator Module Guide Rails 41 2 3 1 4 6 7 5 8 hxn215 eps Figure 21 Remove B Series Valve...
Страница 53: ...Pressure Controller Calibrator Power Supply and Fan 47 3 1 2 4 hxn221 eps Figure 25 Remove Power Supply...
Страница 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Страница 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...
Страница 138: ...6270A Service Manual 132 9 2 5 18 7 9 8 2x 9 5x 3 4x 4 3x 7 2x 11 12 13 6 10 hxn509 eps Figure 56 PMM CAL KIT 20M Parts...