Appendix
93
100
90
80
70
60
50
40
30
20
10
0
R
e
la
tiv
e
h
u
mi
d
ity
%
R
H
10
20
30
40
50
60
70
80
90
100
85
”C
Temperature ”C
95%RH
: LH control range
: LHL, LHU control range
45%RH
• Performance values are given for +23
°C
ambient
temperature and no specimens inside cabinet.
• LH-113 is without a dehumidifying-cooler.
Therefore, temperature-humidity control range,
and especially the low temperature range, will
vary according to installation environmental
conditions (ventilation, fluctuation in room
temperature/humidity, etc.).
• Performance figures are based on JTM K 01-1998
(Humidity chambers-test and indications method
for performance by Japan Testing Machinery
Association).