Chapter 3 Installation
28
To Check Water Level in the Humidifying Tray Water Level Regulator
Water level in the humidifying tray will depend on the water level in the
humidifying tray water level regulator.
The air circulator in the water circuit (heat exhaust) compartment
panel is dangerous. Before opening the compartment panel, set the
main power switch in the OFF position.
1. Detach the water circuit (heat exhaust) compartment panel.
2. Check water level in the humidifying tray water level regulator is
above the WATER LEVEL FOR HUMIDIFIER mark.
7mm
WATER LEVEL FOR
HUMIDIFIER mark
Nut
Humidifying tray
water level regulator
Humidifying tray
Water level sensor
Position adjusting screw
Fig. 3.6 Water level check
3. Reattach the water circuit (heat exhaust) compartment panel.
If water level in the humidifying tray water level regulator is low or
high, regulate it as follows.
1. Loosen the position adjusting screw and raise/lower the
humidifying tray water level regulator.
2. If water level cannot be regulated in step 1, loosen the nut and
turn the water level sensor clockwise (seen from above) to lower
water level, or counter-clockwise to raise it. Then, retighten the
nut.
3. Reattach the water circuit (heat exhaust) compartment panel and
set the main power switch in the ON position.
4. Press the POWER
key.
5. Run the cabinet at 30
and 40%RH.
6. Check water is the proper level.
If water level is okay, the cabinet is ready for use. If not, set the
main power switch in the OFF position and repeat steps 1 ~ 6.
Note
Procedure
Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com