Chapter 6 Checks and Maintenance
78
4. Set the overheat protector about 5°C lower than cabinet temperature.
If the overheat protector is working properly, a buzzer will sound and
the Alarm screen will be appear on the display when temperature
inside the cabinet reaches the overheat protector setting.
If the buzzer doesn’t sound, something is wrong with the equipment.
Contact the place of purchase or ESPEC CORP.
5. To silence the buzzer, press either the
or
keys.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com