Chapter 7 Troubleshooting
85
Clearing Alarms
This section explains how to deal with trouble that trips one of the
cabinet’s self-checks. The buzzer can be silenced by pressing any of the
keys, but follow the below procedure as best possible.
When a 1st Degree Alarm Occurs
1. Press either the
or
keys to silence the buzzer.
2. Check the alarm code
on the display.
3. If the cabinet is running, press the OPER./STOP key to stop the
cabinet.
4. If control power is ON, press the POWER key to shut OFF power
to the instrumentation.
5. Set the main power switch in the OFF position.
6. Remedy the trouble as explained in the Alarm Tables on the following
pages.
7. To resume operation, set the main power switch in the ON position,
press the POWER key and then press the OPER./STOP key
in that order.
When a 2nd Degree Alarm Occurs
When the cause of the trouble persists
1. Check the alarm code
on the display.
2. Press either the
or
keys to silence the buzzer.
3. Remedy the trouble as explained in the Alarm Tables on the following
pages. When the cause of the trouble has been cleared, normal
control is automatically restored. The alarm
code is cleared.
Procedure
Procedure