Chapter 4 Operation
41
3. Upper deviation limit temperature alarm
Stands for
“
Upper Deviation Limit Temperature Alarm
”
Currently set upper deviation limit temperature
:
Changes the value.
:
Selects the digit.
ENT.
:
Enter changes.
* Setting range : 5.0
≦
Upper deviation limit temperature
≦
50.0
○
ALARM SET
Press the
NEXT
key and move to the next setting.
4. Absolute high limit humidity alarm (Not necessary for LU cabinet)
Stands for
“
Absolute High Limit Humidity Alarm
”
Absolute high limit humidity
:
Changes the value.
:
Selects the digit.
ENT.
:
Enter changes.
* Setting range : Target humidity
≦
Absolute high limit humidity
≦
100%RH
○
ALARM SET
Press the
NEXT
key and move to the next setting.
5. Absolute low limit humidity alarm (Not necessary for LU cabinet)
Stands for
Absolute Low Limit Humidity Alarm
Absolute low limit humidity
Changes the value.
Selects the digit.
ENT.
Enter changes.
* Setting range : 0%RH
Absolute low limit humidity
Target
humidity
ALARM SET
Press the
NEXT
key and return to the absolute high limit temperature
alarm setting.
1. Absolute high limit temperature alarm