Chapter 7 Troubleshooting
88
Ty
pe
Displayed alarm
code/channel
code (option)
Trouble Cause
Remedial
action
2nd
23
The humidity inside the
cabinet has dropped below
the absolute low limit. The
heater has been stopped and
refrigerator capacity
decreased to the minimum
(turned OFF in some cases)
until humidity returns within
range.
The absolute low limit
is set higher than the
target humidity.
Set the absolute low limit about
10%RH lower than the target
humidity. When humidity returns
within range, normal control is
restored automatically. The alarm
can be cleared by pressing the
or
keys.
During humidity testing, the
wet-bulb (measures relative
humidity) rose above the
specified temperature.
*This alarm can be triggered
during operation at low
temperature and high
humidity or when the target
humidity is changed. There is
nothing wrong with the
equipment. Silence the
buzzer and continue testing.
Humidity control will
continue as is, but
humidity control might
destabilize and trigger
a humidity alarm.The
wet-bulb wick (gauze)
inside the cabinet
may be dry.
Stop the cabinet and replace the
wick. After wet-bulb temperature
returns within the specified range,
the alarm can be cleared by
pressing the
or
keys.
2nd
The water tank is empty.
Humidity control will continue,
but it will eventually stop
unless water is replenished.
The water tank is
empty.
Refill the portable (option) water
tank and resume testing. After
water has been sufficiently
supplied to the main water tank,
the alarm can be cleared by
pressing the
or
keys. If
the same alarm occurs again, call
for service.
The instrumentation detected
communication trouble
between the CPU board and
display board. The cabinet
has been stopped.
Communication error
between the CPU
board and display
board
Reactivate the system from the
main power switch (leakage
breaker) and resume testing. If the
same alarm occurs again, call for
service.
The instrumentation detected
system trouble in the display
board. The cabinet has been
stopped.
Display board error
(i.e.: internal memory,
etc.)
Reactivate the system from the
main power switch (leakage
breaker) and resume testing. If the
same alarm occurs again, call for
service.
The instrumentation detected
system trouble in the CPU
board. The cabinet has been
stopped.
Sequence task error
on CPU board
Reactivate the system from the
main power switch (leakage
breaker) and resume testing. If the
same alarm occurs again, call for
service.
The instrumentation detected
system trouble in the CPU
board. The cabinet has been
stopped.
Refrigeration task
error on CPU board
Reactivate the system from the
main power switch (leakage
breaker) and resume testing. If the
same alarm occurs again, call for
service.
1st
99
The instrumentation detected
system trouble in the CPU
board. The cabinet has been
stopped.
Temperature control
task error on CPU
board
Reactivate the system from the
main power switch (leakage
breaker) and resume testing. If the
same alarm occurs again, call for
service.