Chapter 4 Operation
55
13. Number of repeat cycles
Input “2” and press the
ENT.
key. Then, press the
NEXT
key.
Once you set the number of repeat cycles, the program will execute the steps in the repeat range
the number of times set here, increasing by 1 each time one cycle is completed. What this means
is that testing returns to the repeat counter start step after the repeat counter end step is executed
instead of proceeding to the next step in the program. Ultimately, when the number set here is
reached, the program will advance to the next chronologically higher number in the program.
Number of repeat cycles
Change the value.
Selects the digit.
ENT.
Enters the selection.
* The number of repeat cycles can be set between 0 and 99.
* If the number of repeat cycles is set to “0”, the program will
be executed as usual without repeating any specific part.
Stands for “Counter”.
PRGM SET
If set to “0”, go to the end mode setting (step 16 in this procedure).
If set to a number between 1 and 99, go to the repeat counter start step
setting (step 14 in this procedure).
Press the
NEXT
key and move to the next setting.
14. Repeat counter start step
Input “1” and press the
ENT.
key. Then, press the
NEXT
key.
Repeat counter start step
Change the value.
ENT.
Enters the selection.
* The repeat counter start step can be set within the following
range. 1
Repeat counter start step
Number of
registered steps
Stands for “Start”.
PRGM SET
Press the
NEXT
key and move to the next setting.