Chapter 4 Operation
58
4.5 Test Startup / End
Test Startup
Before starting tests, make the preparations and settings described in
sections "4.1 Test Preparations", "4.3 Specimen Safety Device Setup”
and "4.4 Target Temperature/Humidity Setup".
1. Check control power is ON.
2. If wanting to run the cabinet in the constant mode, press the
CONST.OPER./STOP key. To run a program, press the
PRGM OPER./STOP key.
•
If the CONST.OPER./STOP key is pressed while testing is in
progress, testing will stop. Pressing it again after that will resume
testing.
•
While a program is running, the program can be stopped and
resumed by pressing the PRGM OPER./STOP key. However,
when the program is resumed, it starts again from step No. 1.
•
You cannot switch between the program and constant modes while
tests are running. First stop the test and then switch the operation
mode.
Test End
Shut the cabinet down as follows.
1. If the cabinet is running in the constant mode, press the
CONST.OPER./STOP key. If a program is running, press the
PRGM OPER./STOP key.
Testing will stop and the cabinet will assume the standby state.
2. Press the POWER key. The display on the operating panel will go
out.
3. If not planning to use the cabinet for a long period of time, set the
main power switch in the OFF position.
Procedure
Reference
Procedure