Chapter 1 Precautions in Usage
15
CAUTION
•
HOT AIR BLAST WHEN OPEN! Use caution when opening
the door during and shortly after operation at high
temperature.
HOT air is blown from inside the cabinet when opened.
•
HOT ON THE INSIDE! During and shortly after operation
above 55°C, the cabinet is HOT on the inside (specimens,
shelves, door gasket, test area walls).
Direct contact may result in burns. Wear heat resistant gloves.
•
COLD ON THE INSIDE! During and shortly after operation
below 0°C, the cabinet is COLD on the inside (specimens,
shelves, door gasket, test area walls).
Direct contact may result in cold burns. Wear cold resistant gloves.
(Not necessary with LH cabinet)
•
When opening the door during or immediately after
operation, be careful of dewing.
Dew formed inside the test area can drip onto the floor. Be mindful of
dewing also when loading specimens which contain moisture into the
cabinet.
•
DO NOT turn the refrigerator ON within 5 minutes of
turning it OFF, or vice-versa.
Failure to observe the above precautions can seriously shorten
refrigerator service-life. (Not necessary with LH cabinet)
•
DO NOT introduce electrically conductive specimens
which might easily be thrown about by air currents inside
the cabinet.
Specimens of the sort can cause current leaks in the heater if they
infiltrate the air conditioner.
•
Do not stack cabinets on top of one another.
•
Keep objects off the cabinet top.
• Blot up moisture from wet specimens before introducing them into the
cabinet.
• The withstand load of the included shelf is 5 kg evenly distributed. Spread
specimens across the entire shelf without bunching them in any one
place.
Note