Chapter 3 Installation
27
Water Consumption Rate
When the FULL lamp is lit
When the FULL lamp is lit, you can use about 14L of water. Though
consumption rate will vary according to conditions of use, the below
table gives a yardstick of water consumption.
Table 3.1 Water consumption rate
Temperature/Humidity
Water
consumption rate
Continuous hours
(days) of operation
60
95
RH 70
40mL/h
Approx. 200
350 hr
(approx. 8
14 days)
85
95
RH 100
130mL/h
Approx. 100
160 hr
(approx. 5
6 days)
When the EMPTY lamp flashes
When the water tank gets low on water, “AL-26” will flash on the display.
If water level drops to about 3L, the EMPTY lamp will start flashing. In
such case, add water. The cabinet can run for another 20 hours from
when the EMPTY first starts flashing until it shuts down completely.
Checking Water Level in the Humidifying Tray
Check water in the humidifying tray is high enough.
•
Check that the humidifying heater is completely under water and that
water is not overflowing from the humidifying tray. An inappropriate
water level will destabilize temperature and humidity control.
When either the humidifying heater is not completely under water or
water is overflowing from the humidifying tray, it is possible that the
cabinet is not level or that the water level in the humidifying tray water
level regulator is too low/high.
If the cabinet is not level, level it. If water level in the humidifying tray
water level regulator is too low/high, regulate it as explained under
"Reference" on the following page.