Chapter 4 Operation
49
5. Target
temperature
Input “20.0” and press the
ENT.
key. Then, press the
NEXT
key.
Target temperature
Changes the value.
Selects the digit.
ENT.
Enter changes.
Setting range : Absolute low limit temperature
Target
temperature
Absolute high limit temperature
The humidity target appears here.
When humidity control is OFF, “OFF” is displayed.
PRGM SET
%RH
Press the
NEXT
key and move to the next setting.
6. Humidity control ON/OFF (Not necessary for LU cabinet)
Select “ON” and press the
ENT.
key. Then, press the
NEXT
key.
PRGM SET
Humidity control status
ON
Turns humidity control ON. Go to the target
humidity setting (step 7 in this procedure).
OFF
Turns humidity control OFF. Go to the
exposure time setting (step 8 in this
procedure).
ENT.
Enters the selection.
Stands for
Humidity
Press the
NEXT
key and move to the next setting.
7. Target humidity (Not necessary for LU cabinet)
Input “80” and press the
ENT.
key. Then, press the
NEXT
key.
Target humidity
Changes the value.
Selects the digit.
ENT.
Enter changes.
*
Setting range : Absolute low limit humidity
Target
humidity
Absolute high limit humidity
RH
PRGM SET
Press the
NEXT
key and move to the next setting.