Chapter 3 Installation
26
3.5 Water Level Check
(Not necessary with LU cabinet)
Water is supplied from the water tank to the humidifying tray. For proper
cabinet operation, water level in the tray must be checked.
•
Use distilled water of a 0.1 to 10 µS/cm conductivity. Using water
with a high impurity content, such as tap water, will shorten
service-life of the humidifying heater and wet-bulb wick.
•
Do not tilt the tank when full. Water will spill out.
To Fill the Water Tank
The water tank is filled by hand. This water is subsequently supplied to
the humidifying tray.
Press the POWER key and run the cabinet at 30ºC and 40%RH. For
details, see “Chapter 4 Operation”.
1. Prepare a water source a minimum of 500 mm higher than the water
tank nipple on the cabinet rear.
2. Connect the included snap-on water supply/drain hose (
9mm ID) to
the water tank nipple. See “3.3 Drainage Work”.
Water supply/drain hose
Water tank nipple
Water tank
Min. 500 mm
Fig. 3.5 Setup for adding water
3. Start water supply.
When the water tank becomes full (approx. 20L), the FULL lamp
(green) will light up. Stop water supply at that point.
Also, if wanting to leave the cabinet running unmanned for 2 to 3 days,
fill the water tank with distilled water in advance.
Note
Procedure
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com