Chapter 4 Operation
30
2. Slide the shelf onto the shelf brackets.
The hook should be to the rear. When sliding the shelf in, make sure
the shelf and hooks clamp the brackets.
Shelf bracket
Shelf
Hook
Fig. 4.2 How to install shelves
How to Charge Specimens
(Use only when needed.)
Use the specimen power supply control terminal (option) to charge
specimens. The rated electric capacity of these terminals is 250V AC, 1
A. A suitable diameter cable would be 0.3 to 2 mm
2
.
DANGER
•
Use only the specimen power supply control
terminals to apply voltage to specimens.
Using other means to charge specimens can drive up
temperature inside the cabinet as specimens generate heat. This
can damage specimens and, in worse cases, result in fire.
•
Always shut OFF power to specimens at the source
before unplugging the specimen power supply
control terminal plug from the connector.
Disconnecting the plug with power ON can result in electric
shock.
CAUTION
•
Droop cables below and up to specimens as shown
in Fig. 4.3.
If cables are taut, dew can run down the cables and damage
specimens.