Chapter 6 Checks and Maintenance
79
6.3 Maintenance
Condenser Cleaning
(Not necessary for LH cabinet)
Before cleaning the condenser, open and close the door to rid the inner
door of any clinging moisture.
1. Unplug the cabinet from its electrical outlet.
2. Remove the screw from the top of the lower front panel. This will
require a Phillips screwdriver.
3. Lean the panel downward from its top edge and lift up and out at an
angle.
Screw
Lower front panel
Fig. 6.2 Detaching the lower front panel
4. Clean the condenser and nearby areas of dust with a vacuum cleaner
or by other means.
5. Reinstall the lower front panel as before.
Procedure