Chapter 4 Operation
42
4.4 Target Temperature/Humidity Setup
Constant Setup Mode
In the constant mode, you set the temperature and humidity.
Getting constant setup mode
Set the main power switch in the ON position and press the POWER
key.
With the temperature monitor on the display, press the
SET
key 1 time.
The target temperature (step 1 in the below procedure) setting will
appear on the display.
Operation
Selects and inputs settings.
Selects the digit.
ENT.
Enters changes to settings.
* Unless the
ENT.
key is pressed, changes you make to settings
will not be updated.
NEXT
Moves to the next setting.
In this manual, indicates a flashing digit.
1. Target
temperature
Target temperature
Changes the value.
Selects the digit.
ENT.
Enter changes.
* Setting range : Absolute low limit temperature
Target
temperature
Absolute high limit temperature
The last set humidity target appears here.
CONST. SET
%RH
Press the
NEXT
key and move to the next setting.